System pressure below 9 psi (62.1 kPa) due to: 1. Teflon® tape is recommended and must be, applied to the male threads omitting the first thread to eliminate the possibility of tape particles entering the oxygen, Electric shock hazard. Afin d’atteindre nos objectifs de qualité, as required or replace tubing if necessary. If bed pressure is rising slowly, check for occluded filters and severe leaks. the side of the fan is directed away from the compressor and. Disconnect the lines from the power switch and circuit breaker. Kompakter Sauerstoffkonzentrator. 3. CAUTION–Do not apply any force or flex the PC Board when, connecting or disconnecting electronic or pneumatic, components. pressures are uneven, continue testing with step 4. K. Use only DeVilbiss concentrator replacement parts and accessories. If any of these readings are not obtained, check for loose or broken, wires in the harness. Also for: 525ks. 525KS. Read and understand all the information contained in this service manual before attempting to operate or perform any maintenance on the concentrator. Category: Health Care Products, Medical Equipment, Oxygen Equipment, Personal … 2. The cabinet should only be removed by a qualified, H. WARNING: Disconnect the power cord from the wall outlet before attempting repairs on the unit. This machi Replace the front and back cabinets and secure with the, 3. 525DS medical equipment pdf manual download. If this condition is, detected by the OSD electronics at any time during concentrator operation, the green “Normal Oxygen” OSD light will turn off and the, beeping audible alert and blinking red “Service Required” light activate. If shipping damage has occurred, contact the DeVilbiss Customer Service Department at 1-800-338-1988 (814-443-4881) for specific instructions. If voltage is not present, the wire harness is defective. 7. After 20 minutes of operation, check the oxygen concentration, levels. described in the chapter Component Testing, Repair and Replacement. Detach cannula from oxygen delivery tubing. materials that are compatible with oxygen. NOTE–Normal pressures observed depend on altitude and flow rate. iMPorTanT SaFeguarDS The capacitor may hold a charge for several days. The DeVilbiss Oxygen Concentrator patient alert system will detect unit component failure. Oxygen leaving the sieve beds is directed through a check valve to the accumulator tank. This unit … 2. While squeezing the locking tabs on the sides of the switch, push the, 5. The valve will stop for several seconds causing the left bed to pressurize while the right bed. 10.9M . Remove the hose from the outlet side of the final bacteria filter. Electric Shock Hazard. also for: 525ds-q, 525ds, 525 series, 525ks, 525ps, drive 1025 series, drive se-1025-1. DANGER–NO SMOKING ES Guía de instrucciones del concentrador de oxígeno de 5 litros de DeVilbiss® AdveRTeNCIA–Lea la guía de instrucciones antes de poner a funcionar WARNING: Electric Shock Hazard. 4. 9. Compact 525 Concentrator Service Manual View and Download DeVilbiss 525DS service manual online. 2. 525Ds Sauerstoffkonzentratoren Pdf Anleitung Herunterladen. seconds, only the “Power” and “Normal Oxygen” lights will remain on. position on the valve plate and discard the used reed valves. 5-Liter Compact Oxygen Concentrator. Remove the screw holding the compressor reed valves in. J. The air filter and oxygen outlet connector should be cleaned at least once a week by the patient. This rapid cycling, only happens on start-up and is clearly heard as pressure is being quickly exhausted several times through the exhaust muffler that is connected to, 2. Health Care Products DeVilbiss 525DS Service Manual . Do not use in rooms heated by paraffin or portable gas heaters. 0044 EN DeVilbiss® 5 Liter Oxygen Concentrator Instruction Guide WARNING–Read instruction guide before operating this equipment. 6. Remove the hose from each end of the final check valve. NOTe– If an audible low-frequency vibration sound is detected, the unit is not operating properly. Read and understand all the information contained in this service manual before attempting to operate or perform any maintenance on the. The DeVilbiss Oxygen Concentrator uses a double-head, oil-free, compressor. 115 Volt Units– Insert the plug into an electrical outlet. behind the bib, it may be loosened or removed. View and Download DeVilbiss 525 Series instruction manual online. Reconnect hose to the compressor intake fitting. As a result, the unit runs cooler. To replace the molecular sieve beds (Figs. 525ks-lt medical equipment pdf manual download. Insert sockets into connector housing and then reconnect the power. to sale by or on the order of a physician. Replace if damaged. To remove the bib completely (Figure 11): CAUTION–The 525KS concentrators use a GSE compressor with, a 10 mfd capacitor. Install the compressor heads so that the holes in the heads are. 4. A defective final check valve may allow water to be drawn in from the, humidifier bottle when the unit is turned off. outlet fitting. Other Clean and replace cannula/mask and tubing as instructed by manufacturer. 4. PERIGO – PROIBIDO FUMAR 525 Series Serie de 525 Série 525 525 Serie Serie da 525 525 Serie Série 525 0044 … 6. When replacing the capacitor, do not touch the terminals or allow metal, objects to come in contact with the terminals on the capacitor. Service interval may be more or less frequent than stated below provided that the Home Care. oxygen, and delivers concentrated oxygen to the patient through an oxygen port. The oxygen patient should perform the following maintenance: Oxygen Humidifier (reusable bottles only), The patient should clean the humidifier bottle daily. Below are the valve voltage test positions and voltages that should. They are manufactured in Pennsylvania. 5 Liter Compact Concentrator Service and Maintenance Log 5 Liter Oxygen Concentrator Built upon a legacy of reliable, field-proven concentrator models, the DeVilbiss 5 Liter Oxygen Concentrator was updated with enhanced quality, durability, performance and simplicity in mind. Le concentrateur d’oxygène Compact 525 a été conçu et fabriqué pour répondre aux demandes les plus exigeantes des patients et offrir un produit de qualité aux prestataires de services. A short DC voltage signal is again applied to the valve. positive airway pressure device (108 pages), Manual will be automatically added to "My Manuals", Why Your Physician Prescribed Supplemental Oxygen, Health Care Products DeVilbiss 525DS Service Manual, Health Care Products DeVilbiss iFill 535D Service Manual, Health Care Products DeVilbiss DV5HH Instructions Manual, Health Care Products DeVilbiss Intellipap DV5HH Use And Care Instructions Manual, Health Care Products DeVilbiss 8650D Instruction Manual, Health Care Products DeVilbiss iGo 306 SERIES Service Manual, Page 9: Guidance And Manufacturer's Declaration. 4. This will allow access to, CAUTION–The 525DS concentrators use a GSE compressor with, a 22 mfd capacitor. or burning objects or naked flames. La société développe, fabrique et commercialise, à travers un réseau de revendeurs en France et … 525 Serie Medizinische Ausstattung Pdf Anleitung Herunterladen. EN DeVilbiss® 5 Liter Oxygen Concentrator Instruction Guide WARNING– Read instRuction guide befoRe opeRating this equipment. to sale by or on the order of a physician. 5 DeVilbiss 525KS manuals available for free PDF download: Instruction Manual, Service Manual DeVilbiss 525KS Instruction Manual (128 pages) Devilbiss 525KS Manuals | ManualsLib The latest Compact 525 oxygen concentrator is designed and built to meet the high demands of patients and cost challenges faced by During this time, the sieve bed pressures. Compact 525 Concentrator Service Manual View and Download DeVilbiss 525DS service manual online. Wash or replace the cabinet air filter. Related Manuals for DeVilbiss 525KS. Press the power switch to the “On” position. Refer to the Normal Operating Sequence to make sure the unit is cycling properly. pressurize causing the pressure relief valve to release the excess pressure. Should. Our Drive DeVilbiss oxygen concentrator repair service is affordable, fast, and includes free inbound shipping. before putting the concentrator back in service. 2. Check line voltage; use alternate circuit. The valve will stop for approximately a second. View and Download DeVilbiss 525DS service manual online. Voltage is again applied to the valve for a short time. Extra care. (If using a prefill, go to Step 3. Install a new regulator on the accumulator tank and attach the tubing, The printed circuit (PC) board is responsible for monitoring and. CAUTION˜ Federal (U.S.A.) law restricts this device to sale by or on the order of a physician. to sale by or on the order of a physician. The following figures show the exterior of the DeVilbiss 5-Liter Compact Concentrators. The air is drawn into the unit through air filters and into a double-head. Verify that the cooling fan is running; if it is not, determine where you are losing power. 7. For Product. Used DEVILBISS HEALTHCARE 525KS Oxygen Concentrator For Sale - DOTmed Listing #3470567: Used, very good technical condition, U.S.A. production, Power supply: 230 V, … NOTE–Make sure that the sealing caps remain on the new sieve beds. Optional SPA-167-K unrestricted full needle travel end cap for remote fluid control using a … WARNING: Do not use oils, greases, or any petroleum-based solvents/cleaners on or near the unit. An oxygen concentrator is a device that delivers highly concentrated oxygen for therapeutic applications. Protect your investment while saving money when you have your DeVilbiss 525DS 5 liter oxygen concentrator repaired by Altra Service Professionals. Respiratory Technical Service Department at 1-800-338-1988 (814-443-4881). Do not use in rooms heated by paraffin or portable gas heaters. Do not attempt to defeat this safety feature. just below the control knob using a 1/16" Allen bit. 5. Download Owners Manual NEW DevilBiss 525DS 5LPM Oxygen Concentrator The DevilBiss 525 is a small and lightweight 5LPM Oxygen Concentrator that is proudly “MADE IN AMERICA “ by Drive-DevilBiss Inc. 6. Apply, leak test solution to all fittings and hose connections with unit, running. the yellow “Low Oxygen” light will turn off and the red “Service Required” light will turn on. T A B LE O F CON T EN T S. GENERAL INFORMATION Position the unit at least 6 inches (16 cm) from walls, draperies, or any other objects that might prevent the proper flow of air in and out of the. Drive DeVilbiss 525KS service manual, English. 1. naked flames. La société développe, fabrique et commercialise, à travers un réseau de revendeurs en France et … DeVilbiss JGA has long been regarded around the world as the spray operators first choice JGA Conventional Pressure Spray Gun - DeVilbiss View and Download DeVilbiss 525KS-LT instruction manual online. Compressor may now be removed from the compressor housing. Pressure may only reach a certain level and then, Low oxygen concentration levels and accumulator pressures higher than, normal may indicate defective sieve beds. to the section on Normal Operating Sequence. Do not remove cabinet. compact 525 concentrator service manual read pdf compact 525 concentrator service manual compact CO M P ONEN T T E S T ING , RE P AIR , AN D RE P LACE M EN T, The DeVilbiss Oxygen Concentrator is designed for ease of service. 8. 6. DeVilbiss FLG-5 is a full size manual spray gun range to provide user’s with the competitive advantage View & download of more than 452 Devilbiss PDF user manuals, service manuals, operating guides. They are manufactured in Pennsylvania. Position the new beds so that the bed serial number label is at the, The pressure regulator stabilizes the flow of oxygen to the patient and, establishes back pressure on the system. 1. Check for proper placement of or damage to the rubber o-ring on. Unscrew the regulator from the accumulator tank. Position compressor on the base of the unit so that the studs on the. repaired before putting the concentrator back in service. Also for: 525ds-q, 525ds, 525 series, 525ks, 525ps, Drive 1025 series, Drive se-1025-1. • Remove the power cord completely from the strap. WARNING: Oxygen promotes rapid burning. 3. The DeVilbiss Oxygen Concentrator uses a two-prong polarized plug and is doubleinsulated to protect against electric shock. Plug the unit into an electrical outlet, turn the unit “On,” and check the audible and visible alerts. Page 2. service, safety, and repair procedures for the DeVilbiss Oxygen Concentrator. WARNING–Read instruction guide before operating this equipment. If no cleaning instructions. 3. Service Manual. Loosen ladder clamps and remove pressure intake and exhaust hoses. C. Do not place a humidifier with an oxygen patient unless prescribed by a physician and then only a bubble-type humidifier should, D. Do not connect the oxygen concentrator to an electrical outlet controlled by a wall switch; the outlet should be independent of. As one bed is being pressurized, the other bed is quickly depressurized. The stepper motor on the rotary valve has multiple windings so, there are several voltage readings that need to be checked in order, to determine if the problem is being caused by the PC board or the, valve itself. Remove the hose from each end of the filter and discard the filter. Ansicht Und Herunterladen Devilbiss 525 Serie Bedienungsanleitung Online. will result in the compressor running slower or not starting. Attach the 5/32" (4 mm) I.D. enabled. Reconnect hoses to the fittings at compressor exhaust and filter. The patient should follow the instructions supplied by the manufacturer. Mechanical Hazard. 6. Remove the air filter located in the door on the back of the unit. DeVilbiss 525DS Service Manual. until just prior to connecting hoses and tubing. Remove the oxygen outlet connector (if used) from oxygen outlet port. The, capacitor is strapped into a well molded into the bottom of the unit. The compressor is secured to the base with four motor, 2. Disconnect the two wires from the terminals on the capacitor. All leaks should be repaired. The capacitor enables the compressor to start and run by supplying, voltage to the windings of the compressor motor. Reconnect the compressor electrical connector. Caution–Do not apply leak test solution to any part of the rotary valve or the main PC Board assembly. 4940-01-106-1415 = KB-555 CUP IMPORTANT: Read and follow all instructions and SAFETY PRECAUTIONS before using this equipment. c. Connect the analyzer to the unit’s oxygen outlet port and wait until the display stabilizes. Remove the screw that secures the board to the unit and remove. Refer to Specifications for normal, The key to simple troubleshooting is to recognize which type of problem exists and select the most effective approach to solving the problem. ), 2. 4. No adjustment can be made without loosening the set screw. While the design is modern, the 5-liter per minute concentrator features numerous internal components incorporated into the 303 and 515A DeVilbiss concentrators. Cut the nylon cable tie holding the capacitor in place and remove. 525DS medical equipment pdf manual download. The cabinet should only be removed by a qualified DeVilbiss homecare, Disconnect the power cord from the wall outlet before attempting repairs on the unit. When the unit is turned off and oxygen flow stops, the check valve closes. Also for: 525ds, 525ps, 525ks. From the bottom, of the unit, remove the four 10 mm hex nuts that secure the motor, 5. NOTE–A high pressure condition is indicated by the audible (a “popping” sound) release of pressure from a pressure relief valve located on the. under the outer edge of meter and prying upward. This solution acts as a germicidal agent. 1. DC voltage is supplied by the PC, board to the motor windings causing the internal discs to turn and direct, A defective check valve will result in a decrease in oxygen concentration, If the rotary valve does not shift properly the same bed may continue to. Remove the eight screws that hold the compressor heads in place. If wire harness is okay, replace the main PC. NOTE–Be sure to read all of the steps involved before beginning any of the procedures in this manual. DeVilbiss FLG-5 is a full size manual spray gun range to provide user’s with the competitive advantage View & download of more than 452 Devilbiss PDF user manuals, service manuals, operating guides. 50ft Tubing for the DeVilbiss 5 Litre Oxygen Concentrator 50 ft Tubing For DeVilbiss 525KS 5 Litre Oxygen Concentrator 50ft tubing to be used with the Devilbiss 525KS 5L Oxygen Concentrator.Ê Please note that this item is sold as single tubes. Replace as necessary. Attach the hoses to a new check valve. If the physician has prescribed an oxygen humidifier as part of the patient’s therapy, follow these steps: 1. 5. If it is determined that the valve did not shift properly or the bed. On parts that are sold separately, the part number is. CAUTION-Federal (U.S.A.) law restricts this device. Note wire colors and socket locations before removing wires. only “bubble-type” humidifiers. The PC board also activates the electronic alert system. Respiratory Technical Service Department at 1-800-338-1988 (814-443-4881). When the power switch is turned “On,” listen for the audible alert and check to see if. Periodic Homecare Provider Preventative Maintenance, Every DeVilbiss Oxygen Concentrator is tested at the factory. Refer. Fill the humidifier bottle (part #HUM16) with distilled, demineralized, or boiled water. The 525 footprint is small, slim, sleek, attractive, lightweight design. To assure continued trouble-free performance, the following preventative maintenance. High pressures indicate defective sieve beds. However. 5-Liter Oxygen Concentrator. 525KS. 5-Liter Compact Oxygen Concentrator. Check for proper placement of or damage to the gaskets on the. NOTE–High-end sieve bed pressure should not exceed 1/2 PSI (4 kPa) above high-end accumulator pressures. Install new flow meter in bib and reconnect hoses. made inthe usa of u.s. and imported parts dangeR–no smoKing ES Guía de instrucciones del concentrador de oxígeno de 5 litros de DeVilbiss® NOTE–Check for leaks using a certified leak detection solution such as Snoop® or equivalent (must not contain ethylene glycol). b. The, cooling fan is located in the bottom of the unit below the compressor, A defective cooling fan may cause the compressor’s internal thermoprotective device to activate and shut the compressor off. b. The homecare provider is responsible for: • determining the condition of the concentrator operating environment. Auch Für: 525-Serie, 525Ks. Database contains 3 DeVilbiss 525KS-LT Manuals (available for free online viewing or downloading in PDF): Instruction manual, Service manual . Devilbiss 1025DS Oxygen concentrators are the most reliable, efficient, and convenient source of supplemental oxygen available today. motor mounts protrude through the holes in the base of the unit. For more in-depth classroom type training, DeVilbiss holds oxygen concentrator service schools. The compressor reed valves should be flush with the valve plate. before installing a new cup seal and retainer plate. Smaller, diameter tubing or the addition of any other, flow restriction may prevent obtaining the, Check for leaks in all hoses and fittings, Oxygen delivery tubing is kinked or blocked, Allow unit to run for 20 minutes and retry, Pressure relief valve activated “popping” sound, Be sure tabs are pushed completely into place, Line cord not properly installed or defective, Oxygen concentrator circuit breaker activated, Insert plug in receptacle or replace line cord. If unit circuit breaker opens again, check, Main wiring harness disconnected/defective, 2. 3. different types of problems and the approaches for solutions are as follows: Type I—The unit runs but a low pressure and flow or high pressure condition exists. NOTE–If the concentrator has been unused for an extended period, the unit must run several minutes before the power fail alert will activate. Increases in altitude and flow rate will slightly decrease, accumulator pressures. Altra Service. Customer reviews. after the unit is turned off. A worn or defective compressor can be caused by a defective internal, These components are included in the Compressor Rebuild Kit (part, CAUTION–If the unit has been running recently, the compressor. View and Download DeVilbiss 525DS service manual online. Set the flow meter to maximum recommended liter flow and let the unit run for at least 20 minutes. 1. If there is voltage at the compressor connector, then either the. Remove the hose clamps and hose from the bottom of the sieve. Remove reed valve plates. Remove the 1/16" plugged piece of tubing from the top of each sieve, 2. causing it to shift and alternately pressurize the sieve beds. 6. When the unit is turned “On,” all four indicator lights (Power, Service Required, Low Oxygen, and, Normal Oxygen) on the front panel will briefly illuminate and an audible signal will sound (the patient alert system) momentarily. It is preset at 8.5 psi (58.6 kPa). 3. Install new sieve beds in reverse order using new plastic cable ties. 4. Replace back cabinet and secure with the six screws. NOTE–A built-in thermal cutoff switch will shut the compressor off if it, becomes overheated. User Manuals, Guides and Specifications for your DeVilbiss 525KS-LT Health Care Products, Medical Equipment. Make sure the unit is unplugged from the wall outlet. This service manual was designed to provide Sunrise Medical Respiratory Products Division qualified service technicians and homecare providers with the proper main-tenance, service, safety, and repair procedures for the DeVilbiss Oxygen Concentrator. Place valve plates on the compressor so that heads of reed valve. 3. Slowly turn the flow meter knob until the flow meter ball is centered on the line next to the appropriate flow rate. needed for more accurate voltage testing. Remove the tubing from the fittings at the top of each sieve bed. Connect pressure gauges to the test points at the top of the sieve. Reconnect all electrical wires, terminals, connectors and the tubing, A check valve (Figure 6) located between the outlet of each sieve bed and, the accumulator tank allows oxygen to pass through the sieve beds to the, accumulator tank when the bed pressure is greater than the accumulator, tank pressure. test assembly (part #303DZ-637) included in the Service Kit. 10. To clean, these steps should be followed: 1. This check valve allows oxygen to flow only out of the unit. Remove the front cabinet and loosen the bib. The 525DS oxygen concentrator comes complete with smooth flow wheel bearings and easy access handles on both the sides and top of the unit. Check to be sure the intake bacteria filter is in place. 6. NOTE–For normal system pressures refer to Specifications. 6. Low oxygen concentration and high pressures indicate, 3. Danger – no SMoking. Extra care should be taken if it is necessary to operate the unit with the cabinet removed. DeVilbiss® 5 Liter Oxygen Concentrator Instruction Guide. connectors on the valve or on the PC Board using a voltmeter. 4. EN DeVilbiss® 5 Liter Oxygen Concentrator Instruction Guide WARNING– Read instRuction guide befoRe opeRating this equipment. Also for: 525ds-q, 525ks, 525ks-lt, 525ps. Do not overfill. Restricted input or output air passage, 3. CAUTION– Federal (U.S.A.) law restricts this device to sale by or on the order of a physician. Drive DeVilbiss France est un leader sur le marché du matériel médical en France. Remove the hose connected to the bottom of the flow meter. Leak test all connections with a certified leak detection solution such, as Snoop® or equivalent (must not contain ethylene glycol). construction or components may exist between the unit in hand and what is described in this manual. Check the audible alert and indicator lights every two years. The air is drawn into the unit through air filters and into a double-head compressor. components with the correct compressor side. Keep for future reference. Testing for proper voltage is done at the wire harness. The cycling is controlled by the PC board. CAUTION–Information for preventing damage to the product. Wash in a solution of warm water and dishwashing detergent. the valve exhaust port and exit through the exhaust muffler. The following parts are included in the Service Kit: Pressure/Vacuum gauge (part #PVO2D-601) 1, In addition to the Service Kit, an oxygen analyzer (part #O2ANA) is needed to periodically check oxygen concentration levels. 7. A malfunction in the pressure regulator will cause either a loss or, fluctuation in the oxygen flow which will be seen on the flow meter or a. 4. CAUTION– Federal (U.S.A.) law restricts this device to sale by or on the order of a physician. If filters are clean and there are no leaks, then the compressor is. The. If the unit falls within specification, the extended life intake bacteria filter does not need to be replaced between. Follow the Operating Instructions on the next page. 8. The DeVilbiss 525 DS is one of the newest additions of home oxygen concentrators.